Dental X-Ray Unit
Z70
70kVp - 7mA
Fuse
Fuse holder
Mobile Model
Fig.02
Dental X-Ray Unit
Z70
70kVp - 7mA
5. Additional Information’s
Follow the requerements from the table below, when the technician installing the Z70:
5.2. Fuses Replacement
16
When necessary, the main power supply fuses must be changed, according the Fig.02, below.
The fuse characteristics are described at the pag.04, item 2.1.
Replacing the Main Fuses
Electrical Installation Requirements
5.1.
Additional Information’s
Electrical Power Requirements
The system requires a three
-
wire electrician
-
supplied power
supply
.
The three
-
wires provide two power lines (L) Line and
(N) Neutral and a (G) Ground
.
Refer to template for
recommended AC main box loca on
.
* Informa on considered for 120V. For 230V, consult the XZeal.
Line Voltage* 120 VAC
±10%
Main Fuse 10 AMP
Circuit Breaker 15 AMP
Wire length and gauge requirements
Maximum length of wire and minimum gauge wire (AWG) from the
power panel box to the Base unit.
Accomplish with the local electrical regula ons law.
Avoid malfunc on, risks and injuries, using an effec ve ground.
0-50 feet 14 gauge minimum
50-100 feet 12 gauge minimum
100-150 feet 10 gauge minimum
150-250 feet 8 gauge minimum
250-400 feet 6 gauge minimum