Technical data
SensoLyt
®
700 IQ H
30
ba76206e02
12/2020
8
Technical data
8.1
Measuring characteristics
Measuring principle
Potentiometric measurement using a combination electrode;
Integrated microprocessor electronics, shielded 2-wire connection for
power and data transmission.
Measuring range
Resolution
Temperature
measurement
Temperature
compensation
8.2
Application characteristics
Temperature range
Allowed pH range of the
measuring medium
4 ... 12
Pressure resistance
Sensor with connected SACIQ sensor connection cable:
pH
0.00 ... 14.00 pH (depending on the electrode)
ORP
-2000 mV ... +2000 mV
(depending on the electrode)
pH
0.01 pH
ORP
1 mV
Temperature sensor
Integrated NTC
Measuring range
- 5 °C ... + 60 °C (23 ... 140 °F)
Accuracy
± 0.5 K
Resolution
0.1 K
Response time t
99
of the tempera-
ture sensor
< 15 s
In the range 0 °C ... 60 °C (32 ... 140 °F)
Measuring medium
0 °C ... + 60 °C (32 ... 140 °F)
Storage/transport
- 5 °C ... + 65 °C (23 ... 149 °F)
Max. allowed overpressure
with electrode
SensoLyt
®
SEA, DWA, PtA in-
stalled
10
6
Pa (10 bar) *