Commissioning

SensoLyt

®

700 IQ H

16

ba76206e02

12/2020

Temp. adjustment

-1.5 K ... +1.5 K

The temperature compensation enables the 
adjusting of the temperature sensor to a ref-
erence temperature measurement (offset-
ting the zero point by ±1.5 K).
Notes:

Please position the sensor in a container 
with a least 2 l water, because of the ther-
mal capacity of the sensor.

Stirring occasionally, leave the sensor in 
this container for at least 15 minutes, if 
the temperature difference between the 
water and sensor is > 10 K, for at least 
one hour, then carry out the adjustment.

Save and quit

The settings are stored. 
The display switches to the next higher lev-
el.

Quit

The settings are not stored. 
The display switches to the next higher lev-
el.

Setting

Selection/values

Explanation

Содержание YSI ViSolid 700 IQ

Страница 1: ...OPERATING MANUAL ba76206e02 12 2020 SensoLyt 700 IQ H IQ SENSORNET pH ORP SENSOR FOR CLASS I DIVISION 2 APPLICATIONS...

Страница 2: ...12 2020 Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email info ysi com Internet www ysi com Copyright 2020 Xylem Inc For the most recent version of the m...

Страница 3: ...tion use 9 3 Commissioning 10 3 1 Scope of delivery 10 3 2 Installation 10 3 3 Commissioning Getting the sensor ready for measuring 12 3 4 Setting table for the SensoLyt 700 IQ H 14 4 Measurement oper...

Страница 4: ...30 8 3 General data 32 8 4 Instrument safety 33 8 4 1 General instrument safety 33 8 4 2 Hazardous location ratings 33 8 5 Electrical data 33 9 Indexes 34 9 1 Explanation of the messages 34 9 1 1 Erro...

Страница 5: ...ng manual has a modular structure like the IQ SENSORNET itself It consists of a system operating manual and the operating manuals of all the components used Please file this component operating manual...

Страница 6: ...vision 2 hazardous locations In conjunction with the SensoLyt SEA HP SensoLyt DWA and SensoLyt ECA pH combination electrodes as well as the SensoLyt PtA ORP combination electrode the SensoLyt 700 IQ H...

Страница 7: ...2 1 2 Safety information on the product Note all labels information signs and safety symbols on the product 2 1 3 Further documents providing safety information The following documents provide additio...

Страница 8: ...thorized see chapter 8 TECHNICAL DATA Any other use is considered unauthorized 2 2 2 Requirements for safe operation Note the following points for safe operation The product may only be operated accor...

Страница 9: ...t and verify that the sensor is rated for your specific application GEFAHR Explosion hazard The associated apparatus MIQ BB1 YSI part no 207001Y has to be used for connection Refer to the following do...

Страница 10: ...t dry the plug connections dab them dry or blow GEFAHR Explosion hazard Only hazardous location rated sensor models Hazloc Sensor must be used in hazardous locations Verify the rating on the product n...

Страница 11: ...g connections dab them dry or blow them dry using compressed air Connecting the sensor to the sensor connection cable Do not suspend the sensor on the sensor connection cable Use an armature or electr...

Страница 12: ...1 Connecting the sensor A KCI filled plastic cap is mounted on the tip of the combi nation electrode to keep the electrode active during storage or during longer pauses in measuring The cap must be re...

Страница 13: ...Commissioning 13 ba76206e02 12 2020 3 Screw the protective cap off the plug head connector of the electrode 4 Screw the electrode into the plug head socket of the sensor 5 Push the unit into the sens...

Страница 14: ...ant IQ SENSORNET system operating manual ACHTUNG Push the connected electrode into the sensor right up to the stop so that the connection is watertight Leakiness can cause the sensor to be destroyed 6...

Страница 15: ...point calibration with any one buffer solution The nominal value of the buffer solution has to be entered Calibration valid Displays and specifies which calibration data the measured value calculation...

Страница 16: ...ainer with a least 2 l water because of the ther mal capacity of the sensor Stirring occasionally leave the sensor in this container for at least 15 minutes if the temperature difference between the w...

Страница 17: ...conventional 2 point calibration with two different buffers first buffer solution pH 7 0 0 5 second buffer solution arbitrary The calibration procedure CAL CON 1P enables the conventional single poin...

Страница 18: ...starting make sure that the correct calibration pro cedure is set see section 3 4 SETTING TABLE FOR THE SENSOLYT 700 IQ H 1 Switch to the measured value display with M and select the sensor SensoLyt 7...

Страница 19: ...n The values determined for Slope and Asymmetry potential are dis played Calibration is finished Confirm with OK The display returns to the mea sured value display Display Explanation Have buffer pH 7...

Страница 20: ...values determined for Slope and Asymmetry potential are dis played Calibration is finished Confirm with OK The display returns to the mea sured value display Display Explanation Have any buffer solut...

Страница 21: ...ccessfully calibrated End of the CAL_CON_1P calibra tion The values determined for Slope and Asymmetry potential are dis played Calibration is finished Confirm with OK The display returns to the mea s...

Страница 22: ...alibration The SensoLyt 700 IQ H provides a feature with witch you can reacti vate the last valid calibration if necessary Thus you can immediately continue to measure if a calibration failed Reactiva...

Страница 23: ...dirt can get into the contacting surfaces In a corrosive environment close the socket of the sensor connec tion cable while it is dry with the screw on dust cover SACIQ Plug to protect the electrical...

Страница 24: ...ssary to replace an electrode proceed as follows 1 Unscrew the protective hood from the sensor 2 Use the protective hood as a tool to lever out the electrode 3 Carefully pull out the electrode until t...

Страница 25: ...tenance and changing the electrode 25 ba76206e02 12 2020 5 Screw in a new combination electrode 6 Push the unit into the sensor up to the stop 7 For measuring pull the KCI filled plastic cap off the c...

Страница 26: ...sing of the sensor as electronic refuse Combination electrodes If no official regulations apply to the contrary used and defective elec trodes can be treated as household waste 8 Screw the protective...

Страница 27: ...buffer solutions for pH calibration Protective hood Screw on protective dust cover for sensor connection cable Model Order no SensoLyt SEA 109 115Y SensoLyt DWA 109 119Y SensoLyt ECA 109 117Y SensoLyt...

Страница 28: ...Sensor not connected Connect the sensor Incorrect instrument setting Correct the instrument setting Cause Remedy The slope of the electrode is outside the tolerance see sec tion 4 2 5 Condition the e...

Страница 29: ...e Remedy Maintenance condition is active If the maintenance condition was activated manually e g by pressing the C key Switch off the maintenance condition manually in the menu Display Options see IQ...

Страница 30: ...Temperature range Allowed pH range of the measuring medium 4 12 Pressure resistance Sensor with connected SACIQ sensor connection cable pH 0 00 14 00 pH depending on the electrode ORP 2000 mV 2000 mV...

Страница 31: ...led combination elec trode SensoLyt SEA DWA PtA min 10 cm max 100 m with installed combination elec trode SensoLyt ECA min 10 cm max 60 m with installed combination elec trode SensoLyt SEA HP min 10 c...

Страница 32: ...rrosion with chloride concentra tions 500 mg ln Automatic sensor monitoring SensCheck function Function for glass breakage monitoring of the pH electrode SensoLyt 700 IQ H Stationary measurements in w...

Страница 33: ...nsoLyt 700 IQ H is a non incendive electrical equipment for use in Class I Division 2 hazardous locations Hazardous location rating 8 5 Electrical data Applicable norms UL 61010 1 CAN CSA C22 2 61010...

Страница 34: ...K Message code Message text EA1311 Meas range exceeded or undercut Check process Select other meas range EA2311 Sensor temperature too high Check process and application EA3311 Sensor temperature too...

Страница 35: ...eck terminal and module connections Defective component replace components ES1311 Component hardware defective Contact service ESA311 SensCheck pH electrode defective glass broken Replace pH electrode...

Страница 36: ...in the Einstellungen Settings Serviceinfo Liste aller Teilnehmer menu see system operating manual by an automated query from a superordinate process control e g when connected to the Profibus from the...

Страница 37: ...returning a product for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Pr...

Страница 38: ...Contact Information SensoLyt 700 IQ H 38 ba76206e02 12 2020...

Страница 39: ......

Страница 40: ...yze and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with...

Отзывы: