Commissioning
SensoLyt
®
700 IQ H
14
ba76206e02
12/2020
3.4
Setting table for the SensoLyt
®
700 IQ H
Carrying out settings
Using
<S>
, switch from the measured value display to the main menu
of the settings. Then navigate to the setting menu (setting table) of the
sensor.The exact procedure is given in the relevant IQ
S
ENSOR
N
ET
system operating manual.
ACHTUNG
Push the connected electrode into the sensor
right up to the stop so that the connection is
watertight. Leakiness can cause the sensor to be
destroyed.
6
For measuring, pull the KCI-filled plastic cap off the combina-
tion electrode.
7
Screw the protective hood onto the sensor.
8
If required, assign a user-defined name to the sensor (see rel-
evant IQ
S
ENSOR
N
ET
system operating manual).
9
Set up the sensor (see section
10
Calibrate the sensor (see section
4.2)
.