6.
If the calibration is unsuccessful, an error message will display on the screen.
Press the
Cal
key to exit the calibration error message and return to the Run
screen. Refer to the troubleshooting section of the manual for a possible
solution or contact YSI Technical Support.
Taking Measurements
1.
Before taking measurements, be sure the instrument has been calibrated to
ensure the most accurate readings.
2.
Turn the instrument on, install the sensor guard and then insert the sensors into
the sample.
3.
Ensure the conductivity sensor is completely submerged in the sample. The two
holes near the cable should be covered by the sample for accurate conductivity
readings.
4.
Wait for the readings to stabilize.
5.
Highlight
Save
and press
Enter
to store the reading.