background image

4. 

Highlight 1 point and press 

Enter

.

5. 

If necessary, use the up and down arrow keys to adjust the pH buffer value.

6. 

Press 

Enter

 to complete the calibration or press 

Cal

  to cancel. 

7. 

Calibration Successful

’ will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen.

8. 

If the calibration is unsuccessful, an error message will display on the screen. 
Press the 

Cal

 key to exit the calibration error message and return to the Run 

screen. Refer to the troubleshooting section of the manual for a possible 
solution or contact YSI Technical Supportelectrolyte will overflow.

2 or 3-Point Calibration

1. 

Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH 
readings to stabilize. 

2. 

Press and hold 

Cal

  for three seconds. 

3. 

Highlight 

pH

 and press 

Enter

. If pH is not listed as an option, check the System 

Setup menu to ensure pH is enabled in the ISE Sensor Type menu.

4. 

Highlight 

2 point

 or 

3 point

 and press enter.

5. 

If necessary, use the up and down arrow keys to adjust the pH buffer value.

6. 

Press 

Enter

 to continue to second point.

7. 

Rinse the sensor and place it in the second pH buffer (4/4.01 or 10/9.18). If 
necessary, use the up and down arrow keys to adjust the pH buffer value. 

8. 

Wait approximately 30 to 60 seconds for the pH sensor to stabilize and for the 
temperature reading to stabilize. 

9. 

Press 

Enter

 to either complete the 2-point calibration or to continue with the 3rd 

calibration point. Or, press 

Cal

  to cancel.

10. 

If continuing on with a 3rd calibration point, rinse the sensor and place it in the 
third pH buffer (4/4.01 or 10/9.18). If necessary, use the up and down arrow keys 
to adjust the pH buffer value. 

11. 

Wait approximately 30 to 60 seconds for the pH sensor to stabilize and for the 
temperature reading to stabilize.

12. 

Press 

Enter

 to complete the 3-point calibration. Or, press 

Cal

  to cancel. 

13. 

Calibration Successful

’ will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen.

14. 

If the calibration is unsuccessful, an error message will display on the screen. 
Press the 

Cal

 key to exit the calibration error message and return to the Run 

screen.