background image

Conductivity Calibration

The Pro1030 can be calibrated in conductivity, specific conductance or salinity.   
Calibration of any option will automatically calibrate the others. YSI recommends 
calibrating specific conductance for both ease and accuracy. The following procedure 
outlines the specific conductance calibration.

1. 

Fill a clean container with conductivity calibration solution and place the sensor 
into the solution. The solution must cover the holes of the conductivity sensor 
that are closest to the cable. Ensure the entire conductivity sensor is submerged 
in the solution or the instrument will read approximately half the expected 
value. Gently agitate the probe to remove any air bubbles from the conductivity 
sensor.

2. 

Turn the instrument on and allow the conductivity and temperature readings 
to stabilize. Press the 

Cal

 key. Highlight 

Conductivity

 and press 

Enter

. Next, 

highlight 

Sp. Conductance

 and press 

Enter

3. 

Highlight the units you wish to calibrate, either 

uS/cm

 or 

mS/cm

, and press 

Enter

. 1 mS = 1,000 uS. Next, use the up or down arrow key to enter the value 

of the conductivity solution as it is listed for 25°C. Most conductivity solution is 
labeled with a value at 25°C. Depressing either the up or down arrow key for 5 
seconds will move the changing digit one place to the left. 

4. 

Press 

Enter

 to complete the calibration. Or, press 

Cal

 to cancel the calibration 

and return to the Run screen.    

5. 

Calibration Successful

’ will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen.

ORP Calibration

1. 

Place the clean sensor in ORP calibration solution. Wait for the ORP and 
temperature readings to stabilize.

2. 

Press and hold 

Cal

  for three seconds. 

3. 

Highlight 

ORP

 and press 

Enter

. If ORP is not listed as an option, check the 

System Setup menu to ensure ORP is enabled in the ISE Sensor Type menu.

4. 

Use the up and down arrow keys to adjust the ORP calibration solution value. 

5. 

Wait for the temperature reading to stabilize, then press 

Enter

 to complete the 

calibration or press 

Cal

  to cancel. 

6. 

Calibration Successful

’ will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen.

7. 

If the calibration is unsuccessful, an error message will display on the screen. 
Press the 

Cal

 key to exit the calibration error message and return to the Run 

screen. Refer to the troubleshooting section of the manual for a possible 
solution or contact YSI Technical Support.

Отзывы: