Measurement / Operation
UV 70x IQ SAC
26
ba77046e02
05/2017
Zero adjustment
While the
Sensor check (H2O dest.)
just determines the current state of the
sensor, a
Zero adjustment
can be used to compensate for changes of the
sensor. A zero adjustment can be required if, for example, the
Sensor check
(H2O dest.)
is not successful even after the sensor was thoroughly cleaned.
If the
Zero adjustment
is not successful, the sensor is blocked for measurement.
A successful
Zero adjustment
is required for the determination of measured val
-
ues.
Sensor check
(standard)
Use the
Sensor check (standard)
to check the sensor with a standard solution
(see chapter 6 S
). During
the check, the measured value is compared to a reference of the standard solu
-
tion stored in the sensor.
The sensor should always be very thoroughly cleaned prior to a zero
adjustment.