Vertical installation For vertical installation, mount the sensor in a position that allows interfering ele
-
ments in the measuring gap to escape with the current. In the case of vertical
installation, the minimum flow rate is 0.2 m/s so the measuring gap is kept clean
optimally.
Fig. 3-2 Vertical installation
Protection against
shocks
When installing the sensor, always make sure it cannot knock against a wall or
another obstacle. Make sure the safety margins are sufficient (e. g. with sus
-
pended mounting).
Mount the enclosed shock protectors (see section 3.3.2). Their soft rubber com
-
pound absorbs hard impacts.
Fig. 3-3 Sensor with mounted shock protectors
Flow direction
Measuring gap opening