Multi 3630/3620 IDS
Operation
ba77045e01
09/2019
11
4.2
OxiTop
®
set management
4.2.1 Overview
After switching to operation with the OxiTop
®
-IDS measuring heads with
<M_>
,
the display of the meter shows the
OxiTop set management
(overview of all set/
samples started).
Set
A set can include several (max. 12) samples, for which the following joint set-
tings are specified:
Measuring mode
(BOD [
BOD standard
/
BOD special
],
Pressure p
)
Duration
Set name
Further settings depending on the selected
Measuring mode
Sample
To determine the BOD, one sample can be distributed to several (max. 12)
measuring bottles (parallel determination), for which the following common set-
tings are specified:
Sample name
Further settings depending on the selected
Measuring mode
The meter automatically creates an identifiable name for the sample. The name
consists of a letter (A) and an index (-x) that is automatically increased for fur-
ther samples in the same set, e.g. A-1. The automatically created name can be
changed manually (max. 8 characters [numerals / letters])
Meas. head
Each measuring bottle is closed with an OxiTop
®
-IDS measuring head.
The OxiTop
®
-IDS measuring heads are always assigned to a sample by the
meter.
Example:
All samples started at the same time (e.g. on a common stirring
platform) are included in a set with a common name.
The meter automatically creates an identifiable name for the set.
The name consists of the date (format JJMMTT) and an index (-x),
which is automatically incremented for sets started on the same
day, e.g. 190422-1. The automatically created name can be
changed manually (max. 8 characters [numerals / letters])
In the
Measuring mode
Pressure p
, a sample includes only one
measuring bottle.