The MIQ/IDS module connects IDS sensors to the IQ S
ENSOR
N
ET
.
Suitable sensors: see section 9.5 A
9.1
Measuring
Prerequisites
The IQ S
ENSOR
N
ET
system is in operation
(see IQ S
ENSOR
N
ET
system operating manual).
The sensor is connected to the MIQ/IDS module
Preparatory activi-
ties
Perform the following preparatory activities when you want to measure:
Avoid gas bubbles (e.g. air bubbles) in the test sample.
Use suitable vessels for measurement and calibration
(see operating manual of the VisoTurb
®
900-P sensor).
Heed the minimum depth of immersion for the sensor
CAUTION
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective mea
-
sures must be taken (protective clothing, protective goggles,
etc.).
1. Install the sensor in a flow through armature and provide the sample flow.
or
For measurement and calibration in containers:
Prepare the sensor for measurement (see
1.
Fill the test sample into a lightproof measuring beaker up to a level of at
least 6 cm.
2.
When immersing the sensor in the test sample, hold the sensor at an
angle.
3.
For measuring, position the sensor upright.