MIQ/CHV PLUS
Installation
11
ba75539d03
03/2018
Small
, clamping range 4.5 to 10 mm. This cable gland is suitable for all
IQ S
ENSOR
N
ET
sensor cables.
Large
, clamping range 7 to 13 mm. This cable gland is required for cable
sheaths with an outside diameter of more than 10 mm and is screwed into the
enclosure via an extension piece.
General
installation
instructions
Observe the following points when attaching connecting wires to the terminal
strip
Shorten all wires to be used to the length required for the installation
Always fit all the ends of the wires with wire end sleeves before connecting
them to the terminal strip
Any wires that are not used and project into the enclosure must be cut off as
closely as possible to the cable gland.
Screw a small cable gland with sealing ring into each remaining free opening
and close it with a blind plug.
If necessary, you can order more large cable glands in a set of 4
pieces (Model EW/1, Order No. 480 051).
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm