
Indexes
ColorVis 705 IQ
52
ba77260e01
06/2021
9.1.2 Informative messages
Message code
Message text
IA23Ax
Calculation of negative concentrations
* Perform user calibration or check user calibration
IAC541
Zero adjustment was successfully carried out
* Carry out sensor check
* If required, carry out new user calibration
IC3541
The factory zero adjustment was activated. Make sure the sensor op
-
erates correctly.
IC4541
The last valid user zero adjustment was activated. Make sure the sen
-
sor operates correctly.
IC5541
Invalid user zero adjustment was replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful
zero adjustment to make sure the sensor operates correctly.
IC6541
The invalid user zero adjustment was replaced by the factory zero
adjustment. Caution! Wrong measured values possible. Check
whether the sensor operates correctly or carry out a new successful
user zero adjustment.