Commissioning
ColorVis 705 IQ
18
ba77260e01
06/2021
3.4.3 Settings for ColorVis 705 IQ (main sensor)
Measuring cycle A measuring cycle consists of the cleaning procedure, the adjustment time for
the measuring system and the determination of the measured value. The follow
-
ing graphic demonstrates the relevant settings:
figure 3-6 Measuring cycle of the sensor
Carrying out
settings
Using <S>, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu (setting table) of the sensor. The
procedure is described in detail in your IQ S
ENSOR
N
ET
system operating
manual.
Time
Adjustment time
Measurement
Measuring
cycle
Measuring interval
Cleaning
A change of the following settings will take some time (up to a few
minutes) until it can be seen in the measured value display:
Measuring mode
Num. of Color-sensors