CN 800
Storage
ba15356d04
12/2021
21
).
Clean the inside of the combination electrode. To do so, open the
combination electrode as follows:
Opening the
combination
electrode
NOTE
Never pull the connection cable of the electrode.
The cable might be damaged.
Storage
Between two
measurements
Condition the electrode in standard solution.
Overnight
Rinse the combination electrode with deionized water and dry it. Then put on
the storage vessel and store the electrode upright.
Longer than
overnight
Remove the bridge electrolyte, then rinse the combination electrode with
deionized water and dab it dry with a clean paper towel. Store the combi-
nation electrode in the storage vessel in a dry place. For the following
measurement, prepare the combination electrode as described in section
C
Recommended accessories
1
Unscrew the knurled nut from the connection head.
2
Position the connection cable upward in a straight line with the con-
nection head.
3
Unscrew the connection head.
4
Push the connection head and pressure spring over the connection
cable.
5
Push the inner body out of the shaft.
Store the reference electrode according to the instructions in its
operating manual.
Description
Model
Order no.
Bridge electrolyte
ELY/BR/503
106 575
ISA sample conditioning solution for CN
-
mea-
surement
MZ/NH3/CN
150130
Polishing strips (24 pieces)
S/FK
180130
Содержание wtw CN 800 DIN
Страница 2: ...2 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 5: ...BEDIENUNGSANLEITUNG ba15356d04 12 2021 CN 800 DIN CYANID SELEKTIVE ELEKTRODE BEDIENUNGSANLEITUNG...
Страница 6: ...6 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 14: ...14 ba15356d04 12 2021 Technische Daten CN 800...
Страница 15: ...OPERATING MANUAL ba15356d04 12 2021 CN 800 DIN CYANIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Страница 16: ...16 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 24: ...24 ba15356d04 12 2021 Technical data CN 800...
Страница 25: ...MODE D EMPLOI ba15356f04 12 2021 CN 800 DIN LECTRODE S LECTIVE DE CYANURE MODE D EMPLOI...
Страница 26: ...26 ba15356f04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 35: ...MANUAL DE INSTRUCCIONES ba15356s04 12 2021 CN 800 DIN ELECTRODO SELECTIVO PARA EL CIANURO MANUAL DE...
Страница 36: ...36 ba15356s04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 45: ......