CN 800
Technical data
ba15356d04
12/2021
23
Technical data
Measuring ranges
Reproducibility
pH range
Temperature range
Resistance
Length
Diameter
Cable length
Plug
DIN plug
– Conditioning time too short
– Extend the conditioning time
– Standard solutions too old
– Use new standard solutions
– Inner junctions encrusted
– Leave the bridge electrolyte to react on
the inner junctions for some hours until
the crusts have dissolved.
– Electrode defective
– Exchange the electrode
Cause
Remedy
0.2 ... 260 mg/l CN
-
(recommended 0.2 ... 25 mg/l CN
-
)
8 x 10
-6
... 1 x 10
-2
mol/l CN
-
(recommended 8 x 10
-6
... 1 x 10
-3
mol/l
CN
-
)
± 2 %
0 ... 14
(Before measuring, adjust the pH value of the sample to
pH 10 to 14)
0 ... 80 °C
(shortly up to 100 °C)
< 30 M
Ω
153 mm (including 33 mm connection head)
Shaft:
12 mm
Connection head:
16 mm
1 m
Содержание wtw CN 800 DIN
Страница 2: ...2 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 5: ...BEDIENUNGSANLEITUNG ba15356d04 12 2021 CN 800 DIN CYANID SELEKTIVE ELEKTRODE BEDIENUNGSANLEITUNG...
Страница 6: ...6 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 14: ...14 ba15356d04 12 2021 Technische Daten CN 800...
Страница 15: ...OPERATING MANUAL ba15356d04 12 2021 CN 800 DIN CYANIDE SENSITIVE ELECTRODE OPERATING MANUAL...
Страница 16: ...16 ba15356d04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 24: ...24 ba15356d04 12 2021 Technical data CN 800...
Страница 25: ...MODE D EMPLOI ba15356f04 12 2021 CN 800 DIN LECTRODE S LECTIVE DE CYANURE MODE D EMPLOI...
Страница 26: ...26 ba15356f04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 35: ...MANUAL DE INSTRUCCIONES ba15356s04 12 2021 CN 800 DIN ELECTRODO SELECTIVO PARA EL CIANURO MANUAL DE...
Страница 36: ...36 ba15356s04 12 2021 CN 800 Copyright 2021 Xylem Analytics Germany GmbH...
Страница 45: ......