Measurement / Operation
CarboVis 70x IQ TS
24
ba75950e03
05/2017
4
Measurement / Operation
4.1
Determination of measured values
Fig. 4-1 Determination of measured values
During a measurement, the absorbance spectrum of the test sample is recorded
and set against the stored absorbance spectrum of a zero adjustment (measure
-
ment in ultrapure water). The algorithm for the calculation depends on the
selected application and/or the measuring location. The resulting raw value can
be adjusted to an independent reference measurement with the aid of a so-
called user calibration.
Absorbance spectrum
of zero adjustment
Absorbance spectrum
of measurement
Physical sensor, series number xxx
Main
sensor
software
Second. sensor
& virtual sensor
software
Main sensor meas. mode
(algorithm)
User calibration data
of main sensor
Main sensor raw value
Meas. value of main sensor
Secondary sensor
raw value
& virtual sensor
Meas. value of sec. sensor
& virtual sensor
Second. sensor
& virtual sensor
meas. mode (algorithm)
User calibration data
of secondary sensor
& virtual sensor
IQ Sensor Net