185
3.6.4
KF Titration
The method being displayed can now be carried out immediately with <
START
>.
The preconditioning is run first.
The solvent and the titration vessel contain moisture (water) that should not influence the calculation of the result.
The conditioning is run automatically after pressing <
START
> (Fig. 69). The final conditions are the same as the
conditions of the actual sample titration.
Fig. 69
When the final criteria are met, then there is an audible signal and
a message appears (
Fig. 70
The conditioning remains active until the actual titration is started by pressing
<Start>
.
You are prompted immediately to add the sample (Fig. 71).
Fig. 71