15 / 25
7. Platinum combination electrodes with Ag/AgCl reference system with 3 mol/l KCl electrolyte are
recommended for redox potential measurements (e. g. EMC133-K010-U).
The description of the used electrode is listed in the separate instructions for use for the sensor.
Configuration PTM40
8.3
8.3.1
Basic settings
Via (
CONF
) you will go to the menu item „
Configuration
“.
The entire structure of the menu (
CONF
) is displayed graphically in
chapter 3.4.
The following device-relevant basic settings can be configured:
data logger
Data logger (
chapter 5)
temperature
Temperature
Sensortype
Sensortype
general
General (
chapter 4)
8.3.2
Configuration of the fixed temperature
As the pH measured value is always temperature-compensated, the temperature of the measuring and
calibration media must be measured or set up as a fixed temperature.
If the fixed temperature is activated, it is used for temperature compensation.
> The message
„FIX
“ will appear in the top right side of the display.
> Configuration of the fixed temperature:
„
Temperature
“ > „
Fixed temperature
“
>
, then select
activate
/
deactivate
.
„
Temperature
“ > „
Temp. value
“
>
, then select the fixed temperature value.
Via (
INFO
) or (
CAL
) you select the position to change.
Via (
CONF +1
) or (
LOG -1
) you change the position.
(
ON/OFF
) confirms and saves your setting.
(
MEAS
) terminates the entry dialog without transferring the value.
8.3.3
Configuration of the sensor type
3 different sensor types can be connected to the PTM40.
pH
ORP
ISE
Sensor selection:
Select „
Sensor type
“ >, then
Type
.
Via (
CONF +1
) or (
LOG -1
) you selected the desired setting.
(
ON/OFF
) confirms and saves your setting.
(
MEAS