Calibration Guide
Starting the calibration mode
The device is switched off.
Press and hold the MIN/MAX/HOLD
and CLR keys simultaneously.
After approx. 5 sec. The following
appear on the display each for one
second
“USER“ followed by “CAL“
and
“MENU“.
The device is now in the user calibration
mode and displays the current wet val-
ues and the currently active calibration
point.
Using the calibration
• HOLD/MIN/MAX key
select the previous calibration point
• CLR key
selec t the nex t calibration point
• DP key
save the currently selected calibration
value
• ON/OFF button
switches off the device
Instructions de calibrage
Lancement du mode de calibrage
L’appareil est éteint.
Appuyer simultanément sur les
touches MIN/MAX/HOLD et les
maintenir enfoncées.
Au bout de cinq secondes environ,
l’affichage «USER» suivi de «CAL»
et de «MENU» apparaissent sur
l’écran.
L’appareil se trouve alors en mode cali-
brage
et indique le taux actuel d’humid-
ité ainsi que le point de calibrage ac-
tuellement activé, p. ex. «RH11.1».
Utilisation du menu de calibrage
• Touches HOLD/MIN/MAX
Sélection du point de calibrage
précédent
• Touche CLR
Sélection du point de calibrage suivant
• Touche DP
Enregistrement de la valeur de
calibrage actuellement sélectionnée
• Touche ON/OFF
Arrêt de l’appareil