Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
690
26
ba25314e05
09/2020
11
Water the sensor head and electrode unit in deionized water
for at least an hour.
12
Carefully shake off the drops of water.
13
Fill a new WP 600 membrane cap and screw it on (see section
5.3 C
).
14
Connect the sensor to the measuring transmitter again.
15
After approx. 60 minutes, the sensor is ready for operation.
16
Recalibrate the sensor.