Safety instructions

DurOx 325

6

ba25400e04

10/2020

2

Safety instructions

This operating manual contains special instructions that must be followed 
during the operation of the  D.O. sensor.

Always keep this operating manual in the vicinity of the sensor.

Special user

qualifications

The membrane cap of the D.O. sensor is filled with a small amount of an 
alkaline electrolyte solution. All maintenance work that requires dealing with 
the electrolyte solution must only be carried out by persons who know how to 
safely deal with chemicals.

General safety

instructions

The individual chapters of this operating manual use safety labels like the one 
below to indicate danger:

CAUTION

indicates instructions that must be followed precisely in order to avoid 
slight injuries or damage to the instrument or the environment.

3

Commissioning

Scope of delivery

D.O. sensor DurOx 325, filled with electrolyte solution and ready to 
operate

Calibration and storage vessel, OxiCal

®

-D

1 exchange membrane cap, WP-D

Electrolyte solution, ELY/G

Cleaning solution, RL/G

Polishing strip, SF 300

Operating manual

Note

The membrane cap that is mounted on the sensor for delivery serves mainly 
as a transport protection. Depending on the duration of the transport and 
storage period, it may have a shortened operational lifetime. If the measuring 
system cannot be calibrated (error message of the instrument), please 
proceed according to section R

EPLACING

 

THE

 

ELECTROLYTE

 

SOLUTION

 

AND

 

MEMBRANE

 

CAP

.

Preparing the

sensor for

measurement

Connect the sensor to the measuring instrument. The sensor is immediately 
ready to measure. It is not necessary to polarize the sensor.

Содержание 201570

Страница 1: ...OPERATING MANUAL ba25400e04 10 2020 DurOx 325 DISSOLVED OXYGEN SENSOR OPERATING MANUAL...

Страница 2: ...DurOx 325 2 ba25400e04 10 2020 Copyright 2020 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...4 2 Measuring 7 4 3 Storage 7 5 Maintenance cleaning replacement 8 5 1 General maintenance notes 8 5 2 Outside cleaning 8 5 3 Changing the electrolyte and membrane cap 9 5 4 Cleaning the electrodes 1...

Страница 4: ...Contents DurOx 325 4 ba25400e04 10 2020...

Страница 5: ...asurements in rivers lakes and wastewater 1 Connection head 2 Shaft housing 3 Temperature sensor and auxiliary electrode 4 Ventilation area 5 Lead counter electrode anode 6 Insulator 7 Gold working el...

Страница 6: ...lowed precisely in order to avoid slight injuries or damage to the instrument or the environment 3 Commissioning Scope of delivery D O sensor DurOx 325 filled with electrolyte solution and ready to op...

Страница 7: ...asured is sufficient aeration tank water pipe stream Slowly pull the sensor through the water by hand lake container or Use a flow aid e g stirrer 4 3 Storage Always store the sensor in the calibratio...

Страница 8: ...to contact with the skin rinse thoroughly with water and immediately change contaminated clothing If it comes into contact with the eyes rinse thoroughly with water and consult a doctor Note the safet...

Страница 9: ...contaminated the membrane is damaged the electrolyte solution is exhausted Changing the electrolyte and membrane cap Unscrew the membrane cap Caution Electrolyte solution For disposal of the membrane...

Страница 10: ...gertight using a paper towel Excess electrolyte solution is forced out of the ventilation area Check the filling Inspect the face surface No air bubbles may be present within the dashed circle Air bub...

Страница 11: ...ane cap Caution Electrolyte solution For disposal of the membrane head and electrolyte solution see section 5 6 Rinse the sensor head with deionized water Using the rough side of the wet SF 300 polish...

Страница 12: ...he membrane head Additionally you can prevent air bubbles by throwing the first filling away and refilling the membrane cap Rinse the sensor head with electrolyte solution Hold the sensor inclined and...

Страница 13: ...st criterion The sensor is OK if the measuring instrument displays 1 D O saturation after 2 minutes 5 6 Disposal CAUTION The ELY G electrolyte solution irritates the eyes and skin When dealing with th...

Страница 14: ...sor Replace and refill the membrane cap see section 5 3 Return the sensor The sensor cannot be calibrated Contaminated membrane head Electrolyte depleted Hole in the membrane Membrane cap not screwed...

Страница 15: ...ctrode Lead Shaft housing Connection head Membrane head Cable screw joint POM Membrane FEP Sensor head Epoxy PEEK Insulator PEEK Thermistor housing VA steel 1 4571 Seals FPM Viton Connection cable Len...

Страница 16: ...aracteristics when delivered Zero signal 1 of the saturation value Response time of D O measurement at 20 C 68 F t90 90 of the final value display after 25 s t95 95 of the final value display after 40...

Страница 17: ...exchange membrane caps 3 pieces WP3 D 202 740 Electrolyte solution ELY G 205 217 Cleaning solution for lead counter electrode RL G 205 204 Polishing strip SF 300 203 680 Accessory kit comprising 3 ex...

Страница 18: ...Wear parts and accessories DurOx 325 18 ba25400e04 10 2020...

Страница 19: ......

Страница 20: ...uildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications ex...

Отзывы: