DurOx 325
Maintenance, cleaning, replacement
11
ba25400e04
10/2020
5.4
Cleaning the electrodes
CAUTION
Before starting to work with the sensor, please note the G
ENERAL
MAINTENANCE
INSTRUCTIONS
on page 8.
General
information
Cleaning is only required in cases of slopes too low or too high (sensor
cannot be calibrated) that cannot be resolved by changing the membrane
head and electrolyte solution.
Cleaning the electrodes
Unscrew the membrane cap.
Caution: Electrolyte solution!
For disposal of the membrane
head and electrolyte solution,
see section 5.6.
Rinse the sensor head with
deionized water.
Using the rough side of the
wet
SF 300 polishing strip,
polish off any contamination
from the gold working
electrode using light pressure.
Caution: Do not use any
conventional sandpaper or
glass-fiber brushes.
Rinse the sensor head with
deionized water.
Wipe the counter electrode
with a lint-free paper towel
and carefully remove any
loose white deposits.
Immerse the sensor head
including the counter
electrode in RL/G cleaning
solution.
Allow to react for 1 to 3
minutes.