Datasheet
Series 60 VioLab
pH
pH 60 VioLab - PC 60 VioLab
Measuring range
-2…16
Resolution / Accuracy
0.1, 0.01 / +0.02
Recognized calibration points and
buffers
AUTO:
1…3 / USA, NIST
CUS:
2 user values
Buffers indication
Yes
Calibration report
Yes, viewable and printable
Automatic DHS recognition
Yes
Stability filter
Low – Med - High
mV
pH 60 VioLab - PC 60 VioLab
Range / Resolution
Range: -1000 … +1900 / Resolution: 0.1 / 1
ORP
pH 60 VioLab - PC 60 VioLab
Calibration points
1 point / 475 mV
Conductivity
COND 60 VioLab - PC 60 VioLab
Range / Resolution
0,00 – 20,00 – 200,0 –
2000 μ
S /
2,00 – 20,00 – 200,0 mS
Automatic scale
Recognized calibration points and
buffers
1…5 / 84, 147,
1413 μ
S, 12.88, 111.8 mS,
1 user value
Reference temperature
15…30 °C
Temperature coefficient
0,00…10,00 %/°C
TDS
COND 60 VioLab - PC 60 VioLab
Measuring range / TDS Factor
0,1mg/
l
…200 gr/l / 0.40…1.00
Salinity
COND 60 VioLab - PC 60 VioLab
Measuring range
0,1ppm …100ppt
Temperature
pH 60 VioLab - COND 60 VioLab - PC 60 VioLab
Measuring range
-10…110 °C
Resolution / Accuracy
0,1 / ± 0,5°C
Temperature compensation ATC
(NTC30K
Ω
) and MTC
0…100 °C
System
GLP with calibration Timer
Yes
Internal memory
1000 Dates
Display
High definition colour LCD
Brightness and contrast management
Manual
I
P protection
IP 54
Outputs
Micro USB and RS 232
Power supply
5 V power supply
Power supply tolerance
+ 10%
Sound level during standard operation
< 80 dB
Electric voltage
100 … 240 V
Work frequency
50 … 60 Hz
Maximum absorption
200 mA
Environmental operating conditions
0 … +45 °C
Maximum permissible humidity
< 95 % non-condensing
Maximum altitude of use
2000 m