37
Rev. 1.0 31/01/2020
EN
Access this menu to select the information to print in the header of a Data Logger on a printer:
•
SIMPLE:
-default setting- The device model, Serial Number and date and time of the last calibration
are printed in the header.
•
COMPLETE:
In addition to the information of the SIMPLE format, the calibration data are also printed
(available in the "CALIBRATION DATA" setup menus).
Note:
The header is printed in the following cases:
•
When a printout of values for a certain parameter is started for the first time
•
When a new calibration is performed
•
Example of automatic Data Logger mode
Example: automatic pH recording on internal memory every 2 minutes
•
Access the
LOG SETTING P8.0
setup menu.
•
Enter the
LOG TYPE P8.1
menu, press
and move with directional keys to
MINUTES
.
•
Use the directional keys to change the number that flashes on the display. Enter "
2
" and confirm with
.
Go back to measure mode and go to the
pH
screen
.
The icon
is lit in the lower string of the display, which indicates that an automatic frequency Data
Logger has been set. Press
to start recording; the icon
starts flashing, indicating that the
storage is in progress. The number next to the
M+
icon indicates how much data has been saved for
that parameter.
Press again
to end the recording.
Note
: automatic recording is suspended when the measurement parameter is changed.
•
Example of manual Data Logger mode
Example: recording a Conductivity value in manual mode
•
Access the
LOG SETTING P8.
0 setup menu.
•
Enter the
LOG TYPE P8.
1 menu, press key
and move with the directional keys to
MANUAL
.
•
Confirm with
and return to measure mode, go to the
COND
screen
.
Press
to save the value. The number next to the
M+
icon indicates how much data has been saved
for that parameter.
Note
: Manual or automatic saving of a value is confirmed by a sequence of flashes of the green LED
•
Example of printing a value
Example: manual printing of a pH value with calibration data too
•
Access the
LOG SETTING P8.0
setup menu.
•
Enter the
LOG TYPE P8.1
menu, press
and move with the directional keys to
MANUAL
. Confirm
with
.