Operation Status Register
Figure 5 shows the Operation Status Register.
The Questionable Status Register takes the standard SCPI structure. The four individual condition inputs are:
6.0.3.1 Measuring
Whenever the output is activated, the unit will be measuring. Therefore, Output Activated serves as this condition.
6.0.3.2 Output Activated
This is the indication that the output is currently activated.
6.0.3.3 Constant Voltage
The output is in its constant voltage mode, where voltage is held constant at a specified setpoint and current will vary dependent
on load.
6.0.3.4 Constant Current
The output is in its constant current mode, where current is held constant at a specified setpoint and voltage will vary dependent
on load.
6.0.3.5 Constant Power
The output is in its constant power mode, where power is held constant at a specified setpoint and voltage/current will vary
dependent on load.
Figure 5. Operation Status Register