XT2640 Operating Manual
13 July 2016
Page 150 of 187
COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
LOGDELAY?
‐
‐
‐
Responds with the present delay time for data logging as four NR1 fields indicating
the days, hours, minutes and seconds respectively.
LOGFILE?
‐
‐
‐
Responds with the present file settings for data logging –
1
st
field (NR1) indicates if the file is in binary (1) or ASCII (0) format or saved to
internal memory (2)
2
nd
field (NR1) indicates if a header record will be included (1) or not (0)
3
rd
field (NR1) indicates if an existing file will be append to (1) or overwritten (0)
4
th
field (STRING) is the file name (including extension)
LOGINTERVAL?
‐
‐
‐
Responds with the present interval for data logging as four NR1 fields indicating
the hours, minutes, seconds and 0.01seconds respectively.
LOGRUN?
‐
‐
‐
Responds with the present run time for data logging as four NR1 fields indicating
the days, hours, minutes and seconds respectively (manually controlled if all fields
are 0)
MODE?
‐
‐
‐
Responds with a NR1 showing the present overall operating mode for the XT2640.
0: Single VPA mode
1: Multi‐VPA mode
2: Sync‐VPA mode
3: EN61000‐3‐2 mode
5: EN61000‐3‐12 mode
7: SPECTRUM mode
MTRDIGITS?
‐
‐
‐
Responds with NR1 number of displayed digits for motor measurement results.
MTREFFGROUP?
‐
‐
‐
Responds with the NR1 efficiency group setting for motor power –
0: None
1: IN
2: MIDDLE
3: OUT
MTRPERIOD?
‐
‐
‐
Responds with the Nr1 measurement period for motor measurements‐
0: 10ms measurement period
1: 30ms measurement period
2: 100ms measurement period
3: 300ms measurement period
4: 1s measurement period
5: 3s measurement period
6: 10s measurement period
7: 30s measurement period
8: 100s measurement period
9: Measurement period synchronous with VPA1
10: Measurement period synchronous with VPA2
11: Measurement period synchronous with VPA3
MTRPOLES?
‐
‐
‐
Responds with the NR3 setting for the number of motor poles (set with the
MTRPOLES command)
MTRSLIP?
‐
‐
‐
Responds with the NR1 selection controlling the calculation of motor slip‐
0: no motor slip calculation
1: motor slip uses the frequency from VPA1 as the drive frequency
2: motor slip uses the frequency from VPA2 as the drive frequency
3: motor slip uses the frequency from VPA3 as the drive frequency
PERIOD?
v
VDEF
‐
Responds with the NR1 measurement period setting for VPA
v
–
0: VLF
1: LF
2: 10Hz/0.3s
3: 20Hz/0.1s
4: 45Hz/20ms
5: 150Hz/10ms
6: 500Hz/2ms
7: Full synchronization with VPA1
8: Full synchronization with VPA2
RESPONSE?
v
VDEF
‐
Responds with the NR1 measurement response setting for VPA
v
–
0: Fastest
1: Medium
2: Slow
3: Slowest
SCOPEINRUSH?
‐
‐
‐
Responds with a NR1 –
0: BANDWIDTH and A RANGE will not be overridden during a scope data capture.
1: BANDWIDTH and A RANGE will be overridden during a scope data capture.