OMNIHI5
USER MANUAL
46
MNL-0022
11. Electromagnetic Compatibility (EMC)
This equipment generates, uses, and radiates radio frequency energy. The equipment
may cause radio frequency interference to other medical or non-medical devices and to
radio communications. If this equipment is found to cause interference, which can be
determined by turning on and off the equipment, the operator or qualified service
personnel should attempt take following actions:
• Reorient or relocate the affected device;
• Increase the separation between the equipment and the affected device;
• Power the equipment by another source;
• Consult the service engineer for further suggestions.
Caution: it is user
’s responsibility to assure that this equipment and vicinity
equipment comply with the contents of IEC 60601-1-2 4th Edition.
Caution: do not use any device that might send out RF signals, including cell
phone, radio transceiver and radio control products, which might cause operation
parameters beyond the standards. Please shutdown these devices when you are
near the equipment. The clinician has the responsibility to warn user or any
others to comply with this rule.
Caution: manufacturer will not be responsible for any unauthorized actions that
cause interference.
Table 1
Guidance and manufacture
’s declaration -electromagnetic emission
The XFT-2003E Nerve and Muscle Stimulator is intended for use in the electromagnetic
environment specified below. The user should comply with specifications ensuring the device
is only used in an appropriate environment.
Emission test
Compliance
Electromagnetic environment- guidance
RF emissions
CISPR 11
Group 1
This equipment uses RF energy only for its
internal function. Its RF emissions are very
low and are not likely to cause any
interference in nearby electronic.
RF emissions
CISPR 11
Class B
This equipment is suitable for domestic
establishments and those directly
connected to the public low-voltage power
supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complied