Calibration and Troubleshooting
M370046-01
6-19
6
Isolated Resistive Programming of Voltage Calibration
To calibrate the isolated resistive programming of voltage:
1. Disconnect the load and connect the voltmeter to the output.
2. Connect the 4.000 k
Ω
resistor to the isolated connector. One terminal
to resistive programming of voltage (J3.8), and voltage programming
of voltage (J3.3), and the other to common (J3.2).
3. Set the resistive APG level to 4 k
Ω
. The SCPI command is:
SYST:COMM:APR:LEV 4
4. Set input source as Resistive Isolated for Voltage Programming. The
SCPI command is:
SYST:REM:SOUR IAR
5. Turn the main output to On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
6. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:RES:ISOL
7. Press on the terminal keys – or + to reduce or increase the main
output voltage value. Adjust the main output voltage to 100%.
8. Press any key than the + or – keys to exit from interactive calibration
mode.
All manuals and user guides at all-guides.com
Содержание XTR 33-25
Страница 2: ...All manuals and user guides at all guides com...
Страница 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Страница 18: ...xvi All manuals and user guides at all guides com...
Страница 30: ...1 8 All manuals and user guides at all guides com...
Страница 92: ...3 48 All manuals and user guides at all guides com...
Страница 198: ...5 72 All manuals and user guides at all guides com...
Страница 262: ...A 36 All manuals and user guides at all guides com...
Страница 284: ...IX 4 All manuals and user guides at all guides com...
Страница 285: ...All manuals and user guides at all guides com...