Calibration and Troubleshooting
6-14
M370046-01
Non-isolated Voltage Programming of Current Calibration
To calibrate the non-isolated voltage programming of current:
1. Short the main output with a shunt. Connect a multimeter to measure
the voltage across the shunt.
2. Connect the input voltage source to the non-isolated connector,
current programming (J1.10 - J1.7).
3. Set the Voltage APG level to 4 V. The SCPI command is:
SYST:COMM:APR:LEV:CURR 4
4. Set the input source voltage to 4.000 V.
5. Set the input source as Voltage Non-Isolated for Current. The SCPI
command is:
SYST:REM:SOUR:CURR AVOL
6. Turn the main output On by pressing the OUTPUT ON/OFF button
on the front panel or send the command:
:OUTP ON
The OUTPUT ON/OFF button will illuminate.
7. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:CURR
8. Press the terminal keys – or + to reduce or increase the main output
current value. Adjust the main output current to 100%.
9. Press any key other than the + or – keys to exit from interactive
calibration mode.
All manuals and user guides at all-guides.com
Содержание XTR 33-25
Страница 2: ...All manuals and user guides at all guides com...
Страница 6: ...All manuals and user guides at all guides com a l l g u i d e s c o m...
Страница 18: ...xvi All manuals and user guides at all guides com...
Страница 30: ...1 8 All manuals and user guides at all guides com...
Страница 92: ...3 48 All manuals and user guides at all guides com...
Страница 198: ...5 72 All manuals and user guides at all guides com...
Страница 262: ...A 36 All manuals and user guides at all guides com...
Страница 284: ...IX 4 All manuals and user guides at all guides com...
Страница 285: ...All manuals and user guides at all guides com...