Local Operation
Using Over Voltage Protection (OVP)
60
Operating Manual for XHR Series Power Supply
Resetting the
OVP Circuit
To reset the OVP circuit after it activates:
1. Reduce the power supply’s output voltage setting to below the OVP set point.
2. Press the STANDBY switch IN. The S/D (Shutdown) LED on the front panel
will turn on. The OVP LED turns off.
3. Press the STANDBY switch again to return power to the load and resume normal
operation.
OR
1. Reduce the power supply’s output voltage setting to below the OVP set point.
2. Turn the power supply OFF using the power switch, wait until the front panel
displays go out, then turn the supply back ON again.
TM-XROP-01XN.book Page 60 Monday, March 22, 2004 4:10 PM