Configure Other Protection Mechanisms
M370078-01
5-71
5
Protection Mask (Enable Alarms)
The protection mask allows for the different alarms to be masked,
completely disabling them. This means that the SCPI status and
operations registers will not detect the alarms. You will have no way of
knowing the current operation state of the alarm. The protection mask is a
feature that is only accessible using the SCPI command noted below.
Command:
[:]SYSTem[<channel>]:PROTection[:MASK] {?|<mask>}
Where
<mask> is the bit mask for disabling the alarms. See Table 5-19 for
details.
See “Alarms and Errors” on page 3–18.
Table 5-19
Alarms Bit Mask
Bit
Bit Weight
Bit Name
Description
0
1
AC Fail
Not used. Always 1.
1
2
OTP
Over Temperature protection.
2
4
N/A
Not used.
3
8
Fan Stop
Fan on the main board.
4
16
N/A
Not used.
5
32
Interlock
Not used. Always 1.
6
64
OVP
Not used. Always 1.
7
128
OCP
Not used. Always 1.
8
256
Foldback
Foldback protection.
9
512
Shutdown
External Shutdown pin
10
1024
UVP
Under Voltage protection.
Содержание Watt Series Programmable DC Power Supply XG 850
Страница 2: ......
Страница 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Страница 6: ......
Страница 18: ...xvi...
Страница 30: ...1 8...
Страница 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Страница 82: ...Local Operation 3 38 M370078 01 The SCPI command s for these instructions are SYSTem channel s REMote STATe...
Страница 86: ...Local Operation 3 42 M370078 01 Figure 3 5 Split Supply Operation V V Common Power Supply Power Supply Load...
Страница 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Страница 198: ...5 72...
Страница 262: ...A 36...
Страница 284: ...IX 4...
Страница 285: ......