M A 3 S P E C T R O P H O T O M E T E R
17
5.
Hold the instrument steady and press the Measure button or tap the screen to initiate a measurement in
case the instrument is configured in Manual Mode. If the instrument is configured in Automatic Mode, the
measurement will be triggered as soon as (with a one-second delay) the instrument is correctly positioned
on the check zone and all pins are in contact (green indicators on UI). Continue to hold the instrument
steady until the progress bar in the screen reaches 100%.
NOTE
: If an error occurs after the measurement, tap
OK
on the screen and try measuring again.
6.
Continue with the remaining area measurements to complete the job.
7.
After the last measurement is completed, the quality of the measurement is indicated with a 5 point system.
Five points indicates excellent measurement without variation between the readings, 3-4 points indicates
good measurement and 1-2 points indicates an acceptable measurement. Tap the Left (◄) arrow on the
Job data saved screen to return to the Job List screen.
8.
If the job was created on the instrument using the Add New Job feature, you can edit the name if desired.
Click on the Edit icon at the button of the screen and enter a job name using the virtual keyboard. Click the
Checkmark icon to save the job name.
Deleting Jobs
Single Job
1. To delete a single job or job data, tap the job from the Job list. You will need to need to tap the Down (
▼
)
or Up (
▲
) arrow icons if the job you want to delete is not showing up in the list.