C O L O R   I 7   O P E R A T I O N   M A N U A L  

20

 

MAINTAINING YOUR SPECTROPHOTOMETER 

Cleaning the Calibration Tiles 

Ceramic standards (calibration tiles) are widely used in color science as standards of reflectance 
factor. Their principal virtue is the stability of their reflection properties. If they are to serve their 

intended purpose, it is necessary that the surfaces of these tiles be maintained in a stable 
condition. The cleaning of any precision optic risks degrading the surface. Therefore, the need for 
cleaning should be minimized by returning the tile to its storage case or covering it with a 

protective bag when it is not in use. If cleaning is required, the following procedure is 
recommended. 

Materials Required 

Camel's-Hair Brush: Available at most camera, hardware, or art supply stores. 
Dust free tissue: Use a lens tissue, which meets Federal Specification NNN-P-40A, Type I or 
equivalent. 
To remove dust, lint, and invisible gritty particles, proceed as follows: 

1.

 

Lint and other small particles are usually best observed by illuminating the tile with a diffuse 
light source at an angle so that the light does not produce surface glare. Inspect the tile at an 

angle rather than looking directly at the tile’s front surface. 

2.

 

Brush the tile's surface with a camel's-hair brush. 

3.

 

Breathe a light mist of condensed vapor at the center of the tile. 

4.

 

Immediately wipe the tile's surface 

lightly

 with lens tissue. Avoid smearing the tile with 

natural greases and perspiration from the hands by wiping with the untouched center part of 
the lens tissue. 

 

Cleaning the Black Plastic Blocking Panel 

Dust on the black plastic blocking panel can be very tightly bound by static electricity. To remove 

the dust, blow canned air across the surface. 
 

Cleaning the Black Trap 

The black trap should be cleaned from time to time to remove and dust or contamination that 
may collect inside. 

 

1.

 

Remove the base (1) from the trap (2) by turning to the left and lifting upwards. 

 

(1) 

(2) 

Содержание Color i7

Страница 1: ...Color i7 Benchtop Spectrophotometer Operation Manual...

Страница 2: ......

Страница 3: ...rity and become unsafe WARNING This instrument is not for use in explosive environments CAUTION Do not stare into operating lamp IR emitted from this product Instructions for disposal Please dispose o...

Страница 4: ...ll not revive an otherwise expired warranty nor shall the same extend the duration of a warranty Customer shall be responsible for packaging and shipping the defective product to the service center de...

Страница 5: ...aking Transmission Measurements 14 Transmission Measurement Setup 14 Film Measurements 17 Liquid Measurements 17 Initiating Measurements 18 Taking Haze Measurements 18 UV Control 18 Aperture Control 1...

Страница 6: ...afe working environment 2 Review the Environmental Specifications to make sure you are using your spectrophotometer in an environmentally appropriate location Safety first Consult this documentation i...

Страница 7: ...cycle 480 measurements per hour max Spectral Range 360 to 750 nm Wavelength Interval 10 nm Photometric range 0 0 to 200 Photometric resolution 0 001 reflectance Environmental Specifications Electrica...

Страница 8: ...alibration tile Round green performance tile Black trap for calibration Standards tiles cleaner tissues 17mm 10mm 6mm aperture plates White Fluorescent standard in an envelope Warnings document contai...

Страница 9: ...Door Preview Method Computer Monitor Used for Sample Preview This preview method involves installing an external video to USB adapter for using your computer monitor for sample preview This method wa...

Страница 10: ...ling it towards you Place your sample at the view port and close the sample arm 2 Using your fingertips open the door at the pry notches and gently lower the door to the complete open position See the...

Страница 11: ...r This is helpful since you can adjust your sample placement at the view port for targeting purposes using the monitor to guide your adjustment Installing the LCD sample preview monitor 1 Remove the m...

Страница 12: ...black connector on the cable to the monitor power connection at the rear of the instrument NOTE The white and red connectors on the cable are not required for this installation Figure 4 Monitor cable...

Страница 13: ...ion as programmed The Sample Preview button is used with the Computer Monitor Video Preview Standard button Preview button Trial button NetProfiler Green enabled Yellow not enabled Blank not NetProfil...

Страница 14: ...ration needs to be calibrated Transmission Calibration Procedure To calibrate your spectrophotometer in Transmission Measurement Mode you need to first locate the following items in your transmission...

Страница 15: ...sing with too much force and possibly damaging the sample 3 Use the preview monitor your computer screen or the drop down door to view the sample and adjust the sample measurement targeting area 4 Tri...

Страница 16: ...ur selection in the software The Color i7 transmission sample holder is designed to mount inside the transmission compartment It is used to measure thin films at both the sphere total transmission and...

Страница 17: ...M A N U A L 15 Figure 7 Transmission holder at the sphere total measurement Figure 8 Transmission holder at the lens direct measurement LAV MAV 17 mm and SAV measurements Front of instrument LAV ONLY...

Страница 18: ...transmission measurements position the base plate with the clamp mounting holes on the sphere side For direct transmission measurements LAV only position the base plate with the clamp mounting holes o...

Страница 19: ...6 and 7 You may want to temporarily tighten the carriage base thumb screw to hold the clap in place while positioning the sample Slowly release the sample clamp carriage base to secure the sample Liqu...

Страница 20: ...on is complete you may begin to take Haze measurements 5 Load your sample in the Transmission sample holder within the Transmission compartment 6 Select Measure Standard Measure Trial or press the app...

Страница 21: ...d and make sure it is nearby 2 Open the sample arm to the fully open position 3 Open the sample preview door Apply pressure at the back of the plate if necessary to remove it 4 Locate the plate to be...

Страница 22: ...pe I or equivalent To remove dust lint and invisible gritty particles proceed as follows 1 Lint and other small particles are usually best observed by illuminating the tile with a diffuse light source...

Страница 23: ...glass with your fingers 4 Blow short bursts of clean dry air 4 across the both surface of the trap glass and inside of the trap 5 Reinstall the black glass in the trap The bottom edge of the black gla...

Страница 24: ...r industrial applications On Site Training and Consulting for Color Management If your training needs are specific to your workflow do you have several staff members to train or you cannot interrupt p...

Страница 25: ...49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803 2705 European Headquarters X Rite Europe GmbH Althardstrasse 70 8105 Regensdorf Switzerland Phone 41 44 842 24 00 Fax 41 44...

Страница 26: ...vette and holder a transmission sample holder and a storage case Ci7 800 LCD Video Preview Monitor A VPM I57LCD Toggle Pad Subassembly with White Ceramic Backing User installable replacement toggle pa...

Страница 27: ...quarters X Rite Europe GmbH Althardstrasse 70 8105 Regensdorf Switzerland Phone 41 44 842 24 00 Fax 41 44 842 22 22 Asia Pacific Headquarters X Rite Asia Pacific Limited Suite 2801 28th Floor AXA Towe...

Отзывы: