C O L O R   I 7   O P E R A T I O N   M A N U A L  

7

 

Spectrophotometer Setup

 

Please follow the instructions detailed in the Color i7 Installation 

Card found as the first item in your shipping box when you 
opened it. 
 
 
 
 
 
 

Sample Preview Methods

 

The Color i7 spectrophotometer has three possible sample preview methods: 

Sample Door

this choice enables you to open the sample door and look at the sample 

position in the viewport and make manual adjustments if necessary to obtain optimum 

sample alignment. Go to instructions on using the Sample Door Preview Method. 

Computer Monitor Used for Sample Preview

This preview method involves installing 

an external video to USB adapter for using your computer monitor for sample preview. 
This method was installed when your instrument first arrived at your location and was 
part of the instructions detailed on the Color i7 installation card.  Please refer to the 

instructions on this card.  

LCD Sample Preview Monitor

-

 This optional preview method involves installing a 

special, mini monitor that mounts on top of the Ci7. This choice presents a live picture of 
your sample at the view port and projects it onto the mini monitor mounted on top of the 

spectrophotometer.  Go to instructions on using the LCD Sample Preview Monitor Method.  

 

 

 

 

 

 

 

 

 

 

Содержание Color i7

Страница 1: ...Color i7 Benchtop Spectrophotometer Operation Manual...

Страница 2: ......

Страница 3: ...rity and become unsafe WARNING This instrument is not for use in explosive environments CAUTION Do not stare into operating lamp IR emitted from this product Instructions for disposal Please dispose o...

Страница 4: ...ll not revive an otherwise expired warranty nor shall the same extend the duration of a warranty Customer shall be responsible for packaging and shipping the defective product to the service center de...

Страница 5: ...aking Transmission Measurements 14 Transmission Measurement Setup 14 Film Measurements 17 Liquid Measurements 17 Initiating Measurements 18 Taking Haze Measurements 18 UV Control 18 Aperture Control 1...

Страница 6: ...afe working environment 2 Review the Environmental Specifications to make sure you are using your spectrophotometer in an environmentally appropriate location Safety first Consult this documentation i...

Страница 7: ...cycle 480 measurements per hour max Spectral Range 360 to 750 nm Wavelength Interval 10 nm Photometric range 0 0 to 200 Photometric resolution 0 001 reflectance Environmental Specifications Electrica...

Страница 8: ...alibration tile Round green performance tile Black trap for calibration Standards tiles cleaner tissues 17mm 10mm 6mm aperture plates White Fluorescent standard in an envelope Warnings document contai...

Страница 9: ...Door Preview Method Computer Monitor Used for Sample Preview This preview method involves installing an external video to USB adapter for using your computer monitor for sample preview This method wa...

Страница 10: ...ling it towards you Place your sample at the view port and close the sample arm 2 Using your fingertips open the door at the pry notches and gently lower the door to the complete open position See the...

Страница 11: ...r This is helpful since you can adjust your sample placement at the view port for targeting purposes using the monitor to guide your adjustment Installing the LCD sample preview monitor 1 Remove the m...

Страница 12: ...black connector on the cable to the monitor power connection at the rear of the instrument NOTE The white and red connectors on the cable are not required for this installation Figure 4 Monitor cable...

Страница 13: ...ion as programmed The Sample Preview button is used with the Computer Monitor Video Preview Standard button Preview button Trial button NetProfiler Green enabled Yellow not enabled Blank not NetProfil...

Страница 14: ...ration needs to be calibrated Transmission Calibration Procedure To calibrate your spectrophotometer in Transmission Measurement Mode you need to first locate the following items in your transmission...

Страница 15: ...sing with too much force and possibly damaging the sample 3 Use the preview monitor your computer screen or the drop down door to view the sample and adjust the sample measurement targeting area 4 Tri...

Страница 16: ...ur selection in the software The Color i7 transmission sample holder is designed to mount inside the transmission compartment It is used to measure thin films at both the sphere total transmission and...

Страница 17: ...M A N U A L 15 Figure 7 Transmission holder at the sphere total measurement Figure 8 Transmission holder at the lens direct measurement LAV MAV 17 mm and SAV measurements Front of instrument LAV ONLY...

Страница 18: ...transmission measurements position the base plate with the clamp mounting holes on the sphere side For direct transmission measurements LAV only position the base plate with the clamp mounting holes o...

Страница 19: ...6 and 7 You may want to temporarily tighten the carriage base thumb screw to hold the clap in place while positioning the sample Slowly release the sample clamp carriage base to secure the sample Liqu...

Страница 20: ...on is complete you may begin to take Haze measurements 5 Load your sample in the Transmission sample holder within the Transmission compartment 6 Select Measure Standard Measure Trial or press the app...

Страница 21: ...d and make sure it is nearby 2 Open the sample arm to the fully open position 3 Open the sample preview door Apply pressure at the back of the plate if necessary to remove it 4 Locate the plate to be...

Страница 22: ...pe I or equivalent To remove dust lint and invisible gritty particles proceed as follows 1 Lint and other small particles are usually best observed by illuminating the tile with a diffuse light source...

Страница 23: ...glass with your fingers 4 Blow short bursts of clean dry air 4 across the both surface of the trap glass and inside of the trap 5 Reinstall the black glass in the trap The bottom edge of the black gla...

Страница 24: ...r industrial applications On Site Training and Consulting for Color Management If your training needs are specific to your workflow do you have several staff members to train or you cannot interrupt p...

Страница 25: ...49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803 2705 European Headquarters X Rite Europe GmbH Althardstrasse 70 8105 Regensdorf Switzerland Phone 41 44 842 24 00 Fax 41 44...

Страница 26: ...vette and holder a transmission sample holder and a storage case Ci7 800 LCD Video Preview Monitor A VPM I57LCD Toggle Pad Subassembly with White Ceramic Backing User installable replacement toggle pa...

Страница 27: ...quarters X Rite Europe GmbH Althardstrasse 70 8105 Regensdorf Switzerland Phone 41 44 842 24 00 Fax 41 44 842 22 22 Asia Pacific Headquarters X Rite Asia Pacific Limited Suite 2801 28th Floor AXA Towe...

Отзывы: