RapidMatch
TM
GO Spectrophotometer
7
During charging, the instrument switches on and indicates charging with a
“lightning bolt” in the battery icon. The instrument will not power completely
off while plugged into the computer USB port, however, it will go into standby
mode. This is indicated by a blue pulsing LED.
A full charge from empty will take approximately 6 hours connected to a
computer USB port. Alternatively a standard USB charger can be used.
The charging time will not be reduced with a standard USB charger.
A USB charger with at least 500 mA is required. A standard charger with 1 A
will not charge any faster.
Battery Icon Information
This icon at the top of the screen would indicate the
battery is fully charged.
This icon at the top of the screen would indicate the
battery has a sufficient charge for a substantial number
of measurements.
This icon at the top of the screen would indicate the
battery pack is low, but measurements are still possible.
Battery should be charged soon.
The “lightning bolt” in the center of the icon at the top of
the screen indicates the battery is charging.
Battery status is also reported in the Device Information menu item of the
Settings window.
5. Measure Button
The measure button (1) is located on the side of the instrument. This is the
same button that is used to power on and off the instrument. You can also tap
the center of the screen (2) to initiate a measurement.
NOTE
: In case of a firmware crash, this button acts as a RESET button.
Remove the USB cable, press and hold the button for 10 seconds and then
release. The instrument will turn off. Refer to the Troubleshooting section for
additional information.
(1)
(2)
Содержание AMW106