Factory Reset Option
This option is used reset the instrument to its factory default settings. All
stored sample will also be deleted when activated. Select Yes to reset to the
factory default settings or select No exit the setting without resetting.
11. Calibration
The instrument must be calibrated on the white tile every 14 days.
Measurements with expired calibration are still possible, but with a warning at
the beginning of each job.
NOTE:
Make sure to use the Calibration reference supplied with the
instrument for calibrating. Do not substitute this reference with a reference
from another instrument. The serial number of the Calbration reference is
shown in the Setting>Calibration screen.
•
Calibration should be performed at room temperature (23°+/-1°C and
40-60% relative humidity).
•
The white tile in the calibration reference is severely affected by smudge
marks, dust, and finger prints.
Cleaning the Calibration Reference
The white tile in the calibration reference should be cleaned using a mild
soap and warm water solution, thoroughly rinsed with warm water, and
wiped dry with a clean, lint-free cloth. You must let the reference dry
completely before taking a calibration measurement.
•
Do not move instrument while taking a calibration measurement. If
motion is detected by the contact sensors, an error message will be
displayed and calibration aborted.
•
Make sure the USB cable is unplugged from the instrument before
performing a calibration.
1.
Select the Calibration option from the Settings menu.
Note
: If
“Calibration expired” message appears, advance to Step 2.
2.
Open the Calibration reference and lay it flat.