2.11 R
ESET
The +CLINO PLUS+ shows: no reaction to key operation. Or the display is incomprehensive. Probably you found
a faulty key sequence as yet unknown to us, which the processor cannot interpret.
If you wish to return the +CLINO PLUS+ to factory default setting
.
(Cancel all parameters set by operator). Or
you have special reasons (problems using hold function or zero setting), which force you to a reset in order to
continue operation.
Solution:
Depress first the key
“ON/MODE" and keep it pressed while pressing
“ENTER/HOLD" for longer
than 10 second.
Apart from the calibration data all stored settings are removed or are set to factory defaults.
The +CLINO PLUS+ is now set as follows:
Measuring mode
Absolute
Measuring unit
mm/m 2 decimals
Relative base
100 mm and 10 inch
Absolute zero
Lost, chapter 2.5
Relative zero
Lost, chapter 3.2
2.12 D
ATA
T
RANSMISSION
True RS485 – Interface / connections to various WYLER instruments such as e.g. Levelmeter 2000, WYLER T/C
as well as to a RS 232 com port of a PC (Special WYLER connector required).
3. APPLICATIONS
3.1. A
BSOLUTE MEASUREMENT
For accurate measurements it is advisable to reset the absolute zero point of the +CLINO PLUS+ in accordance
with chapter 2.5. prior to first measurement. Thereafter place the +CLINO PLUS+ on the surface to be measured,
the instrument directly displays the inclination with respect to absolute horizontal or vertical.
Inclined to the right / +08°32‘
Declined to the right / -08°32‘
3.2. R
ELATIVE MEASUREMENT
Place the +CLINO PLUS+ on the reference surface and using the key
“ON/MODE" select the mode
"RELATIVE" the selection indicator will then flash. Using the key
“ENTER/HOLD" will put the relative mode into
operation, allow a moment until the +CLINO PLUS+ displays
showing that the reference is set. The
+CLINO PLUS+ will then display the difference of inclination with respect to the pre-set value. To return to
absolute mode use the key
“ON/MODE"