What to do if...
pHotoFlex / pHotoFlex Turb
90
ba75491e05
04/2009
6.2
Photometry
Measuring range
undercut or exceeded
Obviously incorrect
measured values
6.3
pH value / ORP voltage
Measuring range
undercut or exceeded
Cause
Remedy
– Program not suitable
– Select program with suitable
measuring range
– Dilute the sample
Cause
Remedy
– Measurement disturbed by
external light
– Close the external light cover.
– Cell not correctly inserted
– Insert the cell so that it is
positioned on the bottom of
the cell shaft.
– Cell contaminated
– Clean the cell
– Cell shaft contaminated
– Clean the cell shaft
– Dilution set incorrectly
– Set the dilution
– Selected program unsuitable
– Select other program
– Zero measurement incorrect
– Perform zero measurement
– Blank value incorrect
– Remeasure the blank value
Cause
Remedy
Electrode:
– Air bubble in front of the
diaphragm
– Remove air bubble
– Air in the diaphragm
– Extract air or
moisten diaphragm
– Gel electrolyte dried out
– Replace electrode
Test sample
– The pH value lies outside the
measuring range
– not possible
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...