Settings/functions
The settings can be found in the Configuration / pH & ORP menu.
To switch to the Configuration menu, press the <MENU> key.
Menu item
Possible
setting
Description
Measured parameter
pH & ORP
mV
Calibration /
Calibration record
-
Displays the calibration re-
cord of the last calibration.
Calibration /
Cal. type
TEC
NIST/DIN
Buffer sets to be used for
pH calibration.
For details, see
section 4.6.5.
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measured value dis-
play.
Calibration /
Unit for slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of
-59.16 mV/pH (100 x de-
termined slope/Nernst
slope).
Man. temperature
-20 ... +130 °C
Entry of the manually de-
termined temperature. For
measurements without
temperature sensor only.
Temperature unit
°C, °F
Degrees Celsius
Degrees Fahrenheit
Reset
Reset all settings for the
pH & ORP measuring
mode (see section 4.10.3)
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...