Operation
pHotoFlex / pHotoFlex Turb
64
ba75491e05
04/2009
Measuring
Note
To keep the drift as low as possible, the time for aligning the cell while
pressing and holding the <START/ENTER> key is limited to 30 sec-
onds. After this time, the meter automatically starts measuring or cali-
brating.
1
Press the <M> key (long pressure) repeatedly until the Turbid-
ity measuring mode is selected.
2
Rinse out a clean cell with the sample to be measured:
Pour approximately 10 ml sample into the cell. Close the cell
and rotate it several times before throwing the sample away.
3
Repeat the rinsing procedure twice more.
4
Fill the cell with the sample to be measured (approx. 15 ml).
Close the cell with the black light protection cap.
5
Clean the cell (see section 5.2.2).
6
Insert the cell (see section 4.2).
7
Align the cell:
z
Marked cell
– Align the marking on the cell cap with the marking on the
cell shaft.
– Press and for a short time hold the <START/ENTER>
key until the measured value is displayed.
z
Unmarked cell (see page 62)
– Press and hold the <START/ENTER> key.
– Slowly and in small steps turn the cell by one complete
rotation (by 360 °). After each step wait for a short time
until the displayed measured value is stable.
– Turn the cell back to the position with the lowest mea-
sured value.
8
Release the <START/ENTER> key.
Measurement starts. The measured value is displayed.
Turbidity
157.0
FNU
NTU
01.02.05 15:12
Содержание pHotoFlex
Страница 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Страница 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Страница 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Страница 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Страница 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Страница 65: ...pHotoFlex pHotoFlex Turb Operation 65 ba75491e05 04 2009 9 Repeat the steps 2 to 8 for further samples...
Страница 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Страница 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Страница 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...