pHotoFlex
®
STD
Operation
43
ba75958e07
11/2020
The
Timer
function is started in the
Configuration
/
Timer
menu by entering a
time interval.
4.5.11 Measuring diluted samples
If the concentration of a test sample exceeds the measuring range of a meth-
od, you can dilute the sample by a factor 1 ... 99 so that the concentration of
the diluted test sample is within the measuring range of the method (see pho-
tometry analysis manual). Thus a valid measurement is possible.
After entering the factor for the dilution the meter converts the concentration
to that of the undiluted sample.
The display then indicates the measured value of the undiluted sample.
Entering the fac-
tor of the dilution
The entered dilution factor is only valid for the selected program. The dilution
factor is erased if
the photometer is switched off
a different program number is selected
the factor 0 is entered in the
Dilution
menu.
If a dilution factor is active, it is indicated on the display during measurement
1
Select the program for which a dilution factor is to be entered.
2
Open the menu
Configuration
/
Photometry
/
Dilution
.
The current factor of the dilution is displayed.
3
Open the display for the entry of numerals with
<START/ENTER>
.
4
Enter the factor of the dilution with the number keys.
The factor has to be a whole number between 0 ... 99.
5
Confirm the factor with
<START/ENTER>
.
6
Exit the
Dilution
menu with
<ESC>
.
For the following measurements with the selected program, the con-
centration of the undiluted sampled is displayed as the measurement
result.
Water + sample
0 + 1
i
83: A6/25 MC
Dilution
Содержание pHotoFlex STD
Страница 1: ...OPERATING MANUAL ba75958e07 11 2020 pHotoFlex STD LED FILTER PHOTOMETER ...
Страница 2: ...ba75958e07 11 2020 pHotoFlex STD Copyright 2020 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 74: ...Lists pHotoFlex STD 74 ba75958e07 11 2020 ...
Страница 76: ...Index pHotoFlex STD 76 ba75958e07 11 2020 ...
Страница 79: ...pHotoFlex STD ba75958e07 11 2020 79 ...