Operation
pHotoFlex
®
STD
40
ba75958e07
11/2020
4.5.7
Zero adjustment
The zero adjustment, i. e. measuring and storing the absorbance of a cell
filled with water, is necessary after the meter is switched on.
Additionally, we recommend to carry out a zero adjustment if the ambient
temperature has changed.
Only perform the zero adjustment against distilled water in an optically per-
fect cell. The zero adjustment must be performed separately for each cell
type.
1
If necessary, press the
<M>
key several times until the measured pa-
rameter
Concentration
is selected.
2
Press the
<CAL/ZERO>
key.
The menu for adjustment measurements opens up.
3
Using
<
V
> <
W
>
and
<START/ENTER>
, select and start the
Zero
adjust.
function.
The menu-guided zero adjustment starts.
Follow the instructions on the display.
Photometry
\
Absorbance
0.532
16 mm
610 nm
01.12.11 15:12
Zero adjust.
Measure blank value
Delete blank value
Standard adjustment
Photometry \ Adjustment