pHotoFlex
®
STD
Operation
41
ba75958d05
10/2017
If a standard adjustment is stored for a program, the standard adjust-
ment is automatically used for measurement. The usage of the stan-
dard adjustment is documented together with the measured value and
indicated in the measured value display with [Cal].
6
Enter the nominal value of the standard.
Note:
Enter the decimal separator with
<START/ENTER>
.
7
Confirm the entered nominal value with
<START/ENTER>
.
8
Insert a cell with standard (see section 4.2).
9
Start the measurement of the standard with
<START/ENTER>
.
After measuring, the result of the standard adjustment is dis-
played and stored.
As the result, the measured value and the adjustment (in %) or
Error
is displayed.
10
Confirm the result with
<START/ENTER>
.
The standard adjustment is completed.
The meter is ready to measure.
or:
Discard the result with
<ESC>
.
Subsequently, carry out a new standard adjustment.
mg/l Cu
_
Enter nominal value
i
304: Cu-1 TP
Cell
i
= 10 mm
i
0.600 mg/l Cu:
2.000 (93.2%)
Accept
Measure standard
Содержание pHotoFlex STD
Страница 1: ...OPERATING MANUAL ba75958d05 10 2017 pHotoFlex STD LED FILTER PHOTOMETER ...
Страница 2: ...ba75958d05 10 2017 pHotoFlex STD Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pHotoFlex STD 6 ba75958d05 10 2017 ...
Страница 16: ...Commissioning pHotoFlex STD 16 ba75958d05 10 2017 ...
Страница 62: ...Operation pHotoFlex STD 62 ba75958d05 10 2017 ...
Страница 74: ...Accessories and options pHotoFlex STD 74 ba75958d05 10 2017 ...
Страница 78: ...Index pHotoFlex STD 78 ba75958d05 10 2017 Z Zero adjustment 43 ...
Страница 81: ...pHotoFlex STD ba75958d05 10 2017 81 ...