Standard adjustment (user calibration)
With some of the programs (methods) for concentration measurement,
you can optimize the calibration curve stored in the meter with the aid
of the
Standard adjustment
function.
A standard adjustment is only valid if the deviation compared to the
original calibration is no more than 30%.
A standard adjustment is always stored for the program that is pres-
ently called up. A standard adjustment is only deleted if
a new standard adjustment is carried out
the standard adjustment is erased manually
the meter is reset to delivery status (
Reset
function)
Standard adjustment
The following measurement of the standard applies only to the selected
program.
If data of a standard adjustment are already available, the data of the
last standard adjustment are displayed. Here you can also erase the
data of an active standard adjustment.
1
If necessary, press the
<M>
key several times until the mea-
sured parameter
Concentration
is selected.
2
Select a program with
<PROG>
as necessary.
3
Using
<CAL/ZERO>
, open the
Photometry \ Adjustment
menu.
4
Open the
Standard adjustment
menu with
<
> <
>
and
<START/ENTER>
.
5
Using
<
> <
>
and
<START/ENTER>
, select and start the
Measure standard
function.
The menu-guided standard measurement begins.
Follow the instructions on the display.
Zero adjust.
Measure blank value
Delete blank value
Standard adjustment
Photometry \ Adjustment
Содержание pHotoFlex STD
Страница 1: ...OPERATING MANUAL ba75958d05 10 2017 pHotoFlex STD LED FILTER PHOTOMETER ...
Страница 2: ...ba75958d05 10 2017 pHotoFlex STD Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pHotoFlex STD 6 ba75958d05 10 2017 ...
Страница 16: ...Commissioning pHotoFlex STD 16 ba75958d05 10 2017 ...
Страница 62: ...Operation pHotoFlex STD 62 ba75958d05 10 2017 ...
Страница 74: ...Accessories and options pHotoFlex STD 74 ba75958d05 10 2017 ...
Страница 78: ...Index pHotoFlex STD 78 ba75958d05 10 2017 Z Zero adjustment 43 ...
Страница 81: ...pHotoFlex STD ba75958d05 10 2017 81 ...