Operation
pHotoFlex
®
STD
38
ba75958d05
10/2017
4.5.4
Blank value (reagent blank value)
A blank value is required for every concentration measurement.
For some programs (methods) for concentration measurement, the
blank values are already stored in the meter. They are used automati-
cally. For all other programs, the blank value has to be determined sep-
arately before the first measurement.
Each stored reagent blank value can be replaced by a blank value de-
termined by the user.
You will find more information on blank values in the photometry anal-
ysis manual. A table with the programs and required blank values is in-
cluded in the analysis specifications.
A blank value is always stored for the program that has just been called
up. It remains stored until it is erased (menu item,
Delete blank value
)
or overwritten.
The
Reset
function erases all blank values measured by the user and
restores the blank values stored in the factory.
If a blank value measured by the user is stored for a program, this blank
value is used for measurement. The usage of the blank value mea-
sured by the user is documented and also indicated in the measured
value display.
Measuring the blank
value
The following measurement of the blank value applies only to the se-
lected program.
1
If necessary, press the
<M>
key several times until the mea-
sured parameter
Concentration
is selected.
2
Select a program with
<PROG>
as necessary.
3
Open the adjustment menu with
<CAL/ZERO>
.
4
Using
<
> <
>
and
<START/ENTER>
, select and start the
Measure blank value
function.
The menu-guided blank value measurement begins.
Follow the instructions on the display.
Zero adjust.
Measure blank value
Delete blank value
Standard adjustment
Photometry \ Adjustment
Содержание pHotoFlex STD
Страница 1: ...OPERATING MANUAL ba75958d05 10 2017 pHotoFlex STD LED FILTER PHOTOMETER ...
Страница 2: ...ba75958d05 10 2017 pHotoFlex STD Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Страница 6: ...Contents pHotoFlex STD 6 ba75958d05 10 2017 ...
Страница 16: ...Commissioning pHotoFlex STD 16 ba75958d05 10 2017 ...
Страница 62: ...Operation pHotoFlex STD 62 ba75958d05 10 2017 ...
Страница 74: ...Accessories and options pHotoFlex STD 74 ba75958d05 10 2017 ...
Страница 78: ...Index pHotoFlex STD 78 ba75958d05 10 2017 Z Zero adjustment 43 ...
Страница 81: ...pHotoFlex STD ba75958d05 10 2017 81 ...