Operation
pH 1970i
50
ba75690e01
01/2007
4.7
Reset
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measured parameters (pH
InI) are reset to the default
condition:
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (
InI
) are reset to the delivery
status:
Measuring mode
pH
Asymmetry
0 mV
Slope -59.16
mV/pH
Calibration procedure
AutoCal TEC
Temperature, manual
25 °C
Resolution (pH display)
0.01 (low resolution)
Display during the
pH calibration
Buffer nominal value
Baud rate
4800
Interval 1
(automatic saving)
OFF
Interval 2
(for data transmission)
OFF