pH 1970i
Operation
31
ba75690e01
01/2007
Note
At this point, the ConCal calibration can be broken off with
<M>
. This
corresponds to a
single-point calibration
. For this, the instrument
uses the Nernst slope (-59.2 mV/pH at 25 °C) and the specified asym-
metry of the electrode.
8
To continue the
two-point calibration
, thoroughly rinse the
electrode with deionized water.
9
If necessary, enter the temperature of the second buffer solu-
tion with
<
▲
> <
▼
>
.
10
Immerse the pH electrode into the second buffer solution.
11
Press the
<RUN/ENTER>
key. The second pH value appears
on the display.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with
<
▲
> <
▼
>
.
13
When the measured value is stable, press the
<RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The sensor symbol shows the evaluation of the current calibra-
tion.
14
Press the
<RUN/ENTER>
key. The value of the asymmetry
(mV) appears on the display.
Auto
REL1
TEC
Store
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
ARng
8
Time
Year
Ident
No.
Day.Month
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH
/
mV
m
cm
S/
%
mg/ l
Sal
O
S
pH
s
2 4 8
L O
°
C
TP
Cal
Sal
cm
1/
K
/
%
2 4 8
°
C
TP
Auto
REL1
TEC
Store
AR
1
LoBat
nLF
Lin
Oxi
Tref25
Tref20
ARng
8
Time
Year
Ident
No.
Day.Month
Baud
RCL
REL2
pH
/
m
cm
S/
%
mg/ l
Sal
O
S
pH
5 9 2
Cal
mV
Содержание pH 1970i
Страница 52: ...Operation pH 1970i 52 ba75690e01 01 2007...
Страница 62: ...Technical data pH 1970i 62 ba75690e01 01 2007...
Страница 68: ...Lists pH 1970i 68...
Страница 70: ...Index pH 1970i 70 ba75690e01 01 2007 31...