Application range / Delivery equipment
Instruction manual
26
ba31107de03 07/2017
ba31107de05 07/2017
27
Recommended range of application
OxiTop
measuring systems with inductive stirring system, for respirometric
determination of the biochemical oxygen demand (BOD) according to the self-checking
method.
Delivery equipment
Model
OxiTop
IS 6
OxiTop
IS 12
Order Nr.
208 210
208 211
Inductive stirring system
(Stirring device)
6 OxiTop
instruments
12 OxiTop
instruments
Equipment
3 OxiTop
yellow
3 OxiTop
green
6 OxiTop
yellow
3 OxiTop
green
Accessories:
Accessory set 1
(brown sample bottle,
stirring rods, rubber
quivers)
6 x
12 x
Accessory set 2
(stirring rod remover,
overflow measuring
beaker 164 ml and
432 ml, sodium
hydroxide tablets,
note block)
1 x
1 x
Operation of the Single Measuring System
OxiTop
Measuring principle
BOD measurement with the OxiTop
measuring system is based on pressure
measurement (difference measurement). The measuring is made by pressure
measurement via
piezoresistive electronic pressure sensors
.
With the following functions, the OxiTop
measuring system minimizes the measuring
work and is especially suited to the courses of the respirometric BOD measurement:
•
AUTO TEMP function:
Control of the
temperature adjustment
and
automatic start
of the measurement
(1 hour after starting the OxiTop
measuring system at the earliest, after 3 hours at
the latest). It is not necessary to temper the samples to exactly 20°C before starting
OxiTop
. Samples between 15°C and 20°C can be started immediately as BOD
measurement will not begin until a stable 20°C is reached. This is controlled by
the "AutoTemp" feature.
•
Data logging
Automatic measurement with daily storage of measured value for 5 days:
facilitates measurement without supervising, e. g. over the weekend.
•
Current value
Indication of measured value (0 - 40 digits) corresponds to the digits of the
previous WTW BOD instruments. Conversion of mbar to digits is stored in the
instrument, i. e. evaluation of the measured values remains unchanged.
•
Measuring range reserve
from 40 to 50 digits - for overranging without reset by opening the bottle.
Содержание OxiTop IS 12
Страница 1: ...BEDIENUNGSANLEITUNG OPERATING MANUAL ba31107de05 07 2017 OxiTop IS 6 IS 12 ...
Страница 2: ...2 ba31107de05 07 2017 ...
Страница 14: ...Bedienungsanleitung 14 ba31107de05 07 2017 Inductive Stirring System IS 6 IS 12 ...
Страница 19: ...Technische Daten ba31107de05 07 2017 19 Prüfzertifikat EMV IS 6 ...
Страница 20: ...Technische Daten 20 ba31107de05 07 2017 Prüfzertifikat EMV IS 12 ...
Страница 21: ...Technische Kundenberatung ba31107de05 07 2017 21 ...
Страница 22: ...22 ...
Страница 34: ...Instruction manual 34 ba31107de03 07 2017 Inductive Stirring System IS 6 IS 12 ...
Страница 39: ...Technical data ba31107de03 07 2017 39 Test certificate EMC IS 6 ...
Страница 40: ...Technical data 40 ba31107de03 07 2017 Test certificate EMC IS 12 ...
Страница 41: ...ba31107de03 07 2017 41 ...
Страница 42: ......
Страница 43: ......