BOD
5
determination
Sample preparation
Measurement
28
ba31107de05 07/2017
ba31107de05 07/2017
29
BOD
5
determination
Normally municipal wastewater does not contain toxic or impeding substances.
There are enough nutrient salts and suitable microorganisms. Under these
conditions the
BOD
5
determination
with the OxiTop
measuring system is possible
in the undiluted sample
Required instruments and accessories
OxiTop
measuring system
Inductive stirring system
Incubator thermostatic box (temperature 20°C ± 1K)
Sample bottles brown (nominal volume 510 ml)
Stirring rods
Stirring rod remover
suitable overflow measuring beakers
Rubber quivers
Sodium hydroxide tablets
Selection of sample volume
Estimate the BOD
5
value to be expected for the wastewater sample:
Expected BOD
5
value
80 % of the COD value
Look for corresponding measuring range in the following chart and gather correct
values for sample volume and factor.
Sample volume (ml)
Measuring range (mg/l)
Factor
432
0 - 40
1
365
0 - 80
2
250
0 - 200
5
164
0 - 400
10
97
0 - 800
20
43.5
0 - 2000
50
22.7
0 - 4000
100
Sample preparation
See following WTW application reports:
WTW Application report 895230:
"BOD measurement in household wastewater"
WTW Application report 895231:
"BOD measurement in organically heavily contaminated wastewater"
WTW Application report 895232:
"BOD measurement with presence of impeding or toxic substances"
Measurement
Information
:
To measure the sample volume, usually overflow measuring
beakers or measuring cylinders are used.
Chose the volume according to the measured value expected.
Too large measuring ranges will lead to inaccurate results. To
estimate the measured value you can reckon with approx. 80%
of the CBS value.
•
For
sample preparation and filling of the measuring bottles
see the following
instruction: DIN 38409 part 52: "Measurement of the oxygen consumption" as
well as WTW application reports, chapter "Sample preparation and
measurement".
•
Rinse
measuring bottle with
sample
. Empty thoroughly.
•
Exactly measure
the required oxygen-saturated (thoroughly homogenized)
quantity of the sample
according to information.
•
Put the
magnetic stirring rod
into the bottle.
•
Insert a
rubber quiver
in the neck of the bottle.
•
Put
2 sodium hydroxide tablets
into the rubber quiver with a tweezers
.
(Caution: The tablets must never come into the sample!)
•
Screw OxiTop
directly on sample bottle (tightly close).
Start measurement:
Press S and M simutaneously. (2 seconds) untill the display
shows 00.
Display: Stored values are deleted.
•
Keep the measuring bottle with the OxiTop put on
for 5 days at 20°C
(e.g. in an
incubator). After the meas. temperature has been reached (after 1 hour at the
earliest, after 3 hours at the latest;
AutoTemp function
), the OxiTop
automatically starts the measurement of the oxygen consumption .
•
During the 5 days the sample is continuously stirred. The OxiTop
automatically
stores one value every 24 hours
for 5 days. To have the current value shown
press the M key.
Display current measured value:
Press M until measured value is displayed (1 second).
•
R
eadout of the stored values
after the 5 days have passed.
Содержание OxiTop IS 12
Страница 1: ...BEDIENUNGSANLEITUNG OPERATING MANUAL ba31107de05 07 2017 OxiTop IS 6 IS 12 ...
Страница 2: ...2 ba31107de05 07 2017 ...
Страница 14: ...Bedienungsanleitung 14 ba31107de05 07 2017 Inductive Stirring System IS 6 IS 12 ...
Страница 19: ...Technische Daten ba31107de05 07 2017 19 Prüfzertifikat EMV IS 6 ...
Страница 20: ...Technische Daten 20 ba31107de05 07 2017 Prüfzertifikat EMV IS 12 ...
Страница 21: ...Technische Kundenberatung ba31107de05 07 2017 21 ...
Страница 22: ...22 ...
Страница 34: ...Instruction manual 34 ba31107de03 07 2017 Inductive Stirring System IS 6 IS 12 ...
Страница 39: ...Technical data ba31107de03 07 2017 39 Test certificate EMC IS 6 ...
Страница 40: ...Technical data 40 ba31107de03 07 2017 Test certificate EMC IS 12 ...
Страница 41: ...ba31107de03 07 2017 41 ...
Страница 42: ......
Страница 43: ......